    Operation and Improvements of a Cesium Beam Standard Having 4-Meter Interaction Length

  • Author

    Kartaschoff, Peter

  • Author_Institution
    Swiss Laboratory for Horological Research, Neuchâtel, Switzerland.
  • Issue
    3
  • fYear
    1962
  • Firstpage
    224
  • Lastpage
    230
  • Abstract
    Some factors affecting a long cesium beam standard´s accuracy are discussed and the results of two year´s operation are given. Increasing the interaction length reduces the linewidth and thus the errors in measuring the line center, as long as the signal-to-noise ratio is sufficient. Errors due to nonlinearity of FM used in servo locking a crystal oscillator to the resonance are below 1×10-11; other servo errors are smaller. The described variable C-field measuring technique and demagnetizable Armco iron shields reduce errors due to C-field uncertainty to 1.2×10-12. Cavity phase shift effects are below 1.5×10-11. Standard deviation of single measurements with 4.6 sec averaging time is 2.7×10-11 under good conditions. Considering all effects, including those of microwave power and of cavity tuning, the stated figure of accuracy is ±3×10-11. Two year´s VLF comparisons show agreement to about ±10-10 with respect to USFS.