DocumentCode
948130
Title
Optimal Source Distribution for Hyperthermia at the Center of a Sphere of Muscle Tissue
Author
Rappaport, Carey M. ; Morgenthaler, Frederic R.
Volume
35
Issue
12
fYear
1987
fDate
12/1/1987 12:00:00 AM
Firstpage
1322
Lastpage
1327
Abstract
In noninvasive electromagnetic hyperthermia it is important that applicators deposit as least as much power at the tumor as anywhere else in the intervening tissue. This study determines the ideal penetration limits for heating the center of a volume of muscle tissue according to this constraint. First, it is shown that the most efficiently heated volume is a sphere. Then, using both a diffraction integral formulation and a spherical harmonic modal approach, field distributions within the sphere are derived. It is shown that combining odd harmonics can generate a nearly uniform surface power distribution, and that this distribution is optimal. The power deposition patterns for these distributions for various standard frequencies are presented. It is concluded that 915 MHz is the best frequency for heating spheres less than 9.5 cm in radius.
Keywords
Applicators; Cogeneration; Diffraction; Frequency; Hyperthermia; Muscles; Neoplasms; Power distribution; Power generation; Power system harmonics;
fLanguage
English
Journal_Title
Microwave Theory and Techniques, IEEE Transactions on
Publisher
ieee
ISSN
0018-9480
Type
jour
DOI
10.1109/TMTT.1987.1133855
Filename
1133855