Title
Calibration of a Network Analyzer Without a Thru Connection for Nonlinear and Multiport Measurements
Author
Heuermann, Holger
Author_Institution
Fachhochschule (FH) Aachen, Univ. of Appl. Sci., Aachen
Volume
56
Issue
11
fYear
2008
Firstpage
2505
Lastpage
2510
Abstract
This paper presents the derivation of the novel calibration procedure without thru using the one-port calibration standards short, open, match, a comb generator, and a power sensor at each port. The necessary architecture presented of a modified network analyzer to measure a multiport and nonlinear S -parameter is quite simple. This novel procedure without thru allows to measure nonreciprocal transfer functions without any reciprocal two-port or multiport calibration standard. Measurement results exemplify the very good quality of the novel calibration procedure for linear and nonlinear measurements under real-life conditions.
Keywords
S-parameters; calibration; multiport networks; network analysers; transfer functions; comb generator; multiport measurement; network analyzer calibration; nonlinear S-parameter; nonlinear measurement; nonreciprocal transfer function; power sensor; Mixed-mode-parameters; multiport measurements; nonlinear scattering parameters; scattering parameters;