    Calibration of a Network Analyzer Without a Thru Connection for Nonlinear and Multiport Measurements

  • Author

    Heuermann, Holger

  • Author_Institution
    Fachhochschule (FH) Aachen, Univ. of Appl. Sci., Aachen
  • Volume
    56
  • Issue
    11
  • fYear
    2008
  • Firstpage
    2505
  • Lastpage
    2510
  • Abstract
    This paper presents the derivation of the novel calibration procedure without thru using the one-port calibration standards short, open, match, a comb generator, and a power sensor at each port. The necessary architecture presented of a modified network analyzer to measure a multiport and nonlinear S -parameter is quite simple. This novel procedure without thru allows to measure nonreciprocal transfer functions without any reciprocal two-port or multiport calibration standard. Measurement results exemplify the very good quality of the novel calibration procedure for linear and nonlinear measurements under real-life conditions.
  • Keywords
    S-parameters; calibration; multiport networks; network analysers; transfer functions; comb generator; multiport measurement; network analyzer calibration; nonlinear S-parameter; nonlinear measurement; nonreciprocal transfer function; power sensor; Mixed-mode-parameters; multiport measurements; nonlinear scattering parameters; scattering parameters;