مقايسه و ارزيابي اثر غير يكنواختي ريه و دنده ، دردوز رسيده به ريه در براكي تراپي پستان به وسيله سيستم طراحي درمان وكد MCNPX
عنوان به زبان ديگر
Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX Code
پديد آورندگان
شمسايي، مجتبي نويسنده دانشگاه صنعتي اميركبير,دانشكده مهندسي هسته اي و فيزيك Shamsaei, mojtaba , صالحي يزدي، حسين نويسنده دانشگاه صنعتي اميركبير,دانشكده مهندسي هسته اي و فيزيك salehi yazdi, hossein , ستايشي ، سعيد نويسنده setayeshi, saeed , جابري ، رامين نويسنده Jaberi, Ramin , شعباني، حميدرضا نويسنده دانشگاه علوم پزشكي تهران shabani, hamidreza
اطلاعات موجودي
فصلنامه سال 1389 شماره 28
رتبه نشريه
علمي پژوهشي
تعداد صفحه
7
از صفحه
21
تا صفحه
27
كليدواژه
سرطان پستان , براكي تراپي , سازوكار طراحي درمان , آهنگ دوز بالا , ناهمگني , MCNPX
چكيده لاتين
Introduction: This study investigates to what extent the computed dose received by lung tissue in a commercially available treatment planning system (TPS) for 192Ir high-dose-rate breast brachytherapy is accurate in view of tissue inhomogeneities and presence of ribs. Materials and Methods: A CT scan of the breast was used to construct a patient-equivalent phantom in the clinical treatment planning system. An implant involving 13 plastic catheters and 383 programmed source dwell positions were simulated using the MCNPX code. Results: The results were compared with the corresponding commercial TPS in the form of isodoses and cumulative dose-volume histogram in breast, lung and ribs. The comparison of Monte Carlo results and TPS calculation showed that the isodoses greater than 62% in the breast that were located rather close to the implant or away from the breast curvature surface and lung boundary were in good agreement. TPS calculations, however, overestimated dose in the lung for lower isodose contours and points that were lying near the breast-air boundary and relatively away from the implant.
Discussion and Conclusions: Taking into account the ribs and entering the actual data for breast, rib and lung, revealed an average overestimation of dose in lung in the TPS calculation.