Title of article :
Comparison of reference-free X-ray fluorescence
analysis and X-ray reflectometry for thickness
determination in the nanometer range
Author/Authors :
Michael Kolbe *، نويسنده , , Burkhard Beckhoff، نويسنده , , Michael Krumrey، نويسنده , , Gerhard Ulm، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
X-ray reflectometry is the method of choice to determine the thickness of nanolayered systems with small uncertainties. In
view of known limitations of this method for extremely thin or laterally inhomogeneous layers we compared X-ray reflectometry
with fundamental parameter based X-ray fluorescence analysis using synchrotron radiation in the radiometry laboratory of the
PTB. The results of both methods for a set of sample systems with transition metal layers of various thicknesses deposited on
silicon wafers were compared and showed a good agreement within their respective uncertainties. For the investigation of
layered systems both methods are very appropriate and, in addition, can give complementary information about the layers. Thus,
the density is determined by X-ray reflectometry, and X-ray fluorescence analysis gives information about trace elements within
the layers and the layer homogeneity
Keywords :
Fundamental constants , X-ray reflectometry , X-ray fluorescence
Journal title :
Applied Surface Science
Journal title :
Applied Surface Science