Title of article :
Isotope ratio measurement with a multicollector detector on the cameca IMS nf
Author/Authors :
Jan Lorincik*، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
3
From page :
7301
To page :
7303
Abstract :
The ASU IMS 3f secondary ion microanalyzer has been converted to a multiple ion detection mass spectrometer with double focusing for three neighbouring masses. We present preliminary experimental tests of the modified instrument. For that purpose we chose sulphur isotope ratio measurements in three pyrite samples using a small triple Faraday cup assembly as a multicollector.We achieved high mass resolution for off-axis masses and an accuracy of 32S/34S isotopic ratio of 0.2%.
Keywords :
Isotopic ratio , Multicollector SIMS
Journal title :
Applied Surface Science
Serial Year :
2006
Journal title :
Applied Surface Science
Record number :
1002547
Link To Document :
بازگشت