• Title of article

    Multielement synthetic transmit aperture imaging using temporal encoding

  • Author/Authors

    K.L.، Gammelmark, نويسنده , , J.A.، Jensen, نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -551
  • From page
    552
  • To page
    0
  • Abstract
    A new method to increase the signal-to-noise ratio (SNR) of synthetic transmit aperture imaging is investigated. The approach utilizes multiple elements to emulate a spherical wave, and the conventional short excitation pulse is replaced by a linear frequency-modulated (FM) signal. The approach is evaluated in terms of image quality parameters in comparison to linear array imaging. Field II simulations using an 8.5-MHz linear array transducer with 128 elements show an improvement in lateral resolution of up to 30% and up to 10.75% improvement in contrast resolution for the new approach. Measurements are performed using our experimental multichannel ultrasound scanning system, RASMUS. The designed linear FM signal obtains temporal sidelobes below -55 dB, and SNR investigations show improvements of 4-12 dB. A 30 mm (/spl ap/45%) increase in penetration depth is obtained on a multitarget phantom with 0.5 dB/[cm MHz] attenuation. Furthermore, in vivo images of the abdomen are presented, which demonstrate the clinical application of the new approach.
  • Keywords
    Power-aware
  • Journal title
    IEEE Transactions on Medical Imaging
  • Serial Year
    2003
  • Journal title
    IEEE Transactions on Medical Imaging
  • Record number

    100679