• Title of article

    Rapid registration for wide field of view freehand three-dimensional ultrasound

  • Author/Authors

    A.H.، Gee, نويسنده , , G.M.، Treece, نويسنده , , R.W.، Prager, نويسنده , , C.J.C.، Cash, نويسنده , , L.، Berman, نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -1343
  • From page
    1344
  • To page
    0
  • Abstract
    A freehand scanning protocol is the only way to acquire arbitrary large volumes of three-dimensional ultrasound (US) data. For some applications, multiple freehand sweeps are required to cover the area of interest. Aligning these multiple sweeps is difficult, typically requiring nonrigid image-based registration as well as the readings from the spatial locator attached to the US probe. Conventionally, nonrigid warps are achieved through general elastic spline deformations, which are expensive to compute and difficult to constrain. This paper presents an alternative registration technique, where the warpʹs degrees of freedom are carefully linked to the mechanics of the freehand scanning process. The technique is assessed through an extensive series of in vivo experiments, which reveal a registration precision of a few pixels with comparatively little computational load.
  • Keywords
    Power-aware
  • Journal title
    IEEE Transactions on Medical Imaging
  • Serial Year
    2003
  • Journal title
    IEEE Transactions on Medical Imaging
  • Record number

    100729