Automatic implementation of the method of standard additions in unsegmented flow systems
Author/Authors
Manuel Agudo، نويسنده , , Angel R?os، نويسنده , , Miguel Valc?rcel، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
8
From page
77
To page
84
Abstract
An open/closed flow system allowing variable volumes of both sample and standard calibration solution to be introduced was employed to perform automatic calibration in unsegmented flow systems by using the method of standard additions. The dilution loop thus established was used to homogenize the sample and standard mixture in a constant final volume that was subsequently introduced into the main flow system. The procedure allows direct analysis of liquid samples with no pretreatment and avoids matrix effects. Glucose and fructose were enzymatically determined over the ranges of 0.01–0.1 g l−1 and 0.02–0.1 g l−1 with relative standard deviation of ±0.8% and ± 1.0% for the determination of 0.05 g l−1 (glucose) and 0.07 g l−1 (fructose), respectively. Both analytes were determined in various types of drinks with average errors of 3.8% (glucose) and 4.6% (fructose).
Keywords
Flow injection , Method of standard additions , Fructose , Glucose , calibration
Journal title
Analytica Chimica Acta
Serial Year
1995