Title of article :
Platinum/glassy carbon electrode as detector for liquid chromatographic determination of hydroxyl-containing compounds
Author/Authors :
Innocenzo G. Casella، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
A platinum-modified glassy carbon electrode was investigated as an amperometric sensor for the determination of alcoholic species in acidic medium. The described electrode was tested in liquid chromatography and flow-injection analysis by pulsed amperometric detection. It showed a promising combination among catalytic activity, detection limits and reproducibility for the determination of alcoholic species. The electrode response was stable; over 8 h a signal loss of less than 5% was observed. Linear responses at low concentrations yielded detection limits in the 0.6–1.8 μg ml−1 ranges for the various aliphatic alcohols. Typical relative standard deviations were comprised between 1.2% and 3.8% for all mono-and polyhydric compounds investigated. Linear plots of 1Ip and 1C indicated Langmuir-type adsorption of analytes on the electrode surface. The Pt chemically modified electrode showed a few important differences with the bulk platinum electrode: i.e., a greater extension of linear range for relatively high concentrations and different trend of sensitivity towards the catalytic oxidation of mono-and polyhydric species.
Keywords :
Liquid chromatography , Amperometry , Sensors , Alcohols
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta