Title of article :
Properties of poly-β-aminoanthraquinone modified carbon fiber electrode as a basis for hemoglobin biosensors
Author/Authors :
Huangxian Ju، نويسنده , , Hongbin Sun، نويسنده , , Hongyuan Chen، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Abstract :
An amperometric microsensor for hemoglobin (HB) has been constructed using a carbon fiber electrode modified with poly-(pAAQ). Scanning electron microscopic studies showed that the potential range and time of electrochemical polymerization had strong effects on the morphology of the polymer film at the microelectrode. The stability and the electrochemical properties of the modified electrode were studied using cyclic voltammetry. The pAAQ modified microelectrode showed an excellent electrocatalytic activity for the reduction of HB in a weakly acidic solution. The catalytic current in cyclic voltammetry increased linearly with the HB concentration between 1.0–400 μM. The amperometric response to HB had a linear range of 0.5 μM to 340 μM with a correlation coefficient of 0.990(n = 18). The biosensor was stable for at least one month when it was stored in a phosphate buffer (pH 7.4) below 20 °C. The obtained HB concentration in the whole blood agreed well with the one obtained with the reference instrument.
Keywords :
Voltammetry , Poly-?-aminoanthraquinone , Blood , Hemoglobin , Catalytic methods , Biosensors
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta