• Title of article

    Using the electrochemical quartz crystal microbalance as stripping detector. Application to trace mercury analysis

  • Author/Authors

    Niels Peder Raj Andersen، نويسنده , , Per Holst-Hansen، نويسنده , , Dieter Britz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    4
  • From page
    253
  • To page
    256
  • Abstract
    The electrochemical quartz crystal microbalance (EQCM) was used as a stripping detector for mercury(II) in 0.1 M perchloric acid with a plating time of 15 min. Mercury concentrations down to 1 ppb could be measured. The calibration curve stripping frequency change vs. mercury(II) concentration, showed nonlinearity at mercury(II) concentrations higher than 75 ppb. Kinetic and experimental factors appear to be the cause.
  • Keywords
    Quartz Crystal Microbalance , Mercury
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1996
  • Journal title
    Analytica Chimica Acta
  • Record number

    1023114