Title of article :
Novel potentiometric monitoring combined with coulometric reagent generation
Author/Authors :
Géza Nagy، نويسنده , , Kl?ra T?th، نويسنده , , Zs?fia Fehér، نويسنده , , Jen? Kunovits، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Abstract :
A potentiometric monitor incorporating a flow-through analysis channel has been developed based on the standard addition or subtraction principle. Programmed addition of standard or reagent was carried out at a given point of the flow-through system by current-programmed coulometry. A computer was used for data acquisition, for electric noise rejection, for signal recognition and for calculating the analytical results. The characteristics of the monitor system have been investigated under conditions optimised for the chloride determination, and the performance of it was checked by monitoring the chloride concentration in tap water.
Keywords :
Flow system , Potentiometry , Coulometry
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta