• Title of article

    Highly sensitive chemiluminescence flow injection analysis system using microbial peroxidase and a photodiode detector

  • Author/Authors

    Kenji Hayashi، نويسنده , , Satoshi Sasaki، نويسنده , , Kazunori Ikebukuro، نويسنده , , Isao Karube، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    8
  • From page
    127
  • To page
    134
  • Abstract
    A highly sensitive chemiluminescence flow injection analysis system (CL-FIA) was developed using an Si-photodiode as a detector. The CL-FIA system with peroxidase from Artheromyces ramosus (ARP) showed a higher chemiluminescence intensity than that with the enhanced chemiluminescence of horse radish peroxidase (HRP) and p-iodophenol. The optimum conditions for chemiluminescence using ARP were a carbonate buffer (pH 10.1) containing 17.5 μM luminol in the flow cell. The photocurrent of the photodiode was linearly related to the hydrogen peroxide concentration in a log-log plot from 3 nM to 10 μM under these optimum conditions. The RSD was estimated to be 3.2% for 1 μM of hydrogen peroxide, and the time for one analysis was less than 20 s. A CL-FIA system for hypoxanthine (Hx) measurement was also constructed with this CL-FIA system by addition of a bioreactor column of xanthine oxidase. With its compactness, ease of use, and sensitivity comparable with CL-FIA using a photomultiplier tube, this system has potential applicability for daily measurements in the industrial field.
  • Keywords
    Artheromyces ramosus peroxidase , Hypoxanthine , Flow injection , Chemiluminescence , Photodiode
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1996
  • Journal title
    Analytica Chimica Acta
  • Record number