• Title of article

    Photocured polymers in ion-selective electrode membranes Part 6: Photopolymerized lithium sensitive ion-selective electrodes for flow injection potentiometry

  • Author/Authors

    J.R. Farrell، نويسنده , , P.J. Iles، نويسنده , , T. Dimitrakopoulos، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    6
  • From page
    111
  • To page
    116
  • Abstract
    A photocured lithium ion-selective electrode, based on the ionophore N,N-dicyclohexyl-N′,N′-diisobutyl-cis-cyclohexane-1, 2-dicarboxamide (ETH 1810), was developed and evaluated. The robust nature of the photocured membrane made it ideally suitable for measurements in flow injection potentiometry within a linear range 0.1–10−3 M, detection limit of 5 × 10−4 M, and sample throughput of 150 h−1. The electrode was used successfully to determine lithium levels in pharmaceutical lithium carbonate tablets.
  • Keywords
    Potentiometry , Lithium , Ion-selective electrodes , Flow injection , Photocured membrane
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1996
  • Journal title
    Analytica Chimica Acta
  • Record number