• Title of article

    Flow injection analysis of copper diethyldithiocarbamate in high resistance toluene media using a microelectrode detector

  • Author/Authors

    T.J. Cardwell، نويسنده , , J.H. Santos، نويسنده , , A.M. Bond، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    6
  • From page
    169
  • To page
    174
  • Abstract
    Flow injection (FI) analysis of copper as the copper diethylthiocarbamate complex (Cu(Et2dtc)2) has been performed in a high resistance toluene carrier stream. The method is based on the oxidation of the copper complex at a platinum microelectrode fabricated to fit a Metrohm wall-jet flow cell. Electrodes of sizes 5 to 25 μm radii were tested in toluene-based carriers containing either 0.03 M tetrahexylammonium perchlorate (Hex4NClO4) or 35% acetonitrile. Detection limits (3 : 1 signal to noise ratio) in the range 1 × 10−7 M (5 μm radius electrode) to 5 × 10−8 M (25 μm radius electrode) were achieved, which are comparable to detection limits reported in lower resistance media under similar conditions. The system was applied to the analysis of copper in water samples after extraction of the metal as its diethyldithiocarbamate into toluene and good agreement was found with data obtained by other voltammetric methods.
  • Keywords
    Microelectrodes , Toluene media , copper , Flow injection
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1997
  • Journal title
    Analytica Chimica Acta
  • Record number

    1024428