Title of article
Determination of chlorine dioxide using 4-aminoantipyrine and phenol by flow injection analysis
Author/Authors
Takaho Watanabe، نويسنده , , Toshikazu Ishii، نويسنده , , Yoshihiro Yoshimura، نويسنده , , Hiroyuki Nakazawa، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
6
From page
257
To page
262
Abstract
The determination of chlorine dioxide (ClO2) using 4-aminoantipyrine (4-AA) and phenol by flow injection analysis, was investigated in the presence of various chlorinated species that can occur under normal water-treatment conditions. The colorimetric method with 4-AA and phenol relies on the use of a Tris-HCl buffer. Adjusting the reagent to pH 9.0 with the Tris-HCl buffer gives high sensitivity and is effective in eliminating hypochlorite interference. Furthermore, using purge trap system for pretreatment makes it possible for this method to eliminate interferences from metal ions, as well as other oxychlorinated compounds, such as chlorite and chlorate.
Keywords
4-Aminoantipyrine , Chlorine dioxide , Row injection , Phenol