    Calibration transfer from a scanning near-IR spectrophotometer to a FT-near-IR spectrophotometer

  • Author/Authors

    Jie Lin، نويسنده , , Su-Chin Lo، نويسنده , , Chris W. Brown، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    7
  • From page
    263
  • To page
    269
  • Abstract
    A principal component regression calibration model developed on a scanning near-IR spectrophotometer (with a standard cuvette) was transferred to a Fourier transform near-IR spectrophotometer (with a fiber-optic probe) for the determination of NaCl in aqueous solutions by using piecewise direct standardization method. Only two transfer samples were needed in obtaining a good calibration transfer. Standard error of prediction obtainable from the calibration transfer was 0.034 M, compared to 0.012 M produced by the full set recalibration for NaCl in the concentration range of 0–5 M.
  • Keywords
    Calibration transfer , Principal component regression (PCR) , Near-IR spectroscopy , NaCl solution , Instrument standardization
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1997