• Title of article

    Development of a membrane inlet mass spectrometric method for analysis of air samples

  • Author/Authors

    R.A. Ketola، نويسنده , , M. Sampot and M. Ojala، نويسنده , , H. Sorsa، نويسنده , , T. Kotiaho، نويسنده , , R.K. Kostiainen، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    7
  • From page
    359
  • To page
    365
  • Abstract
    A sheet membrane inlet mass spectrometric (MIMS) method for the on-line analysis of volatile organic compounds (VOCs) in air at low μg/m3 levels was developed. The effects of the thickness of the membrane, the temperature of the membrane inlet and the flow rate of the sample on the responses and response times of some selected VOCs were studied in detail. Under optimised conditions the detection limits for the VOCs studied were 0.5–4 g m−3 and linear dynamic ranges were about four orders of magnitude. Response times of only a few seconds were recorded with a thin (25 μm) membrane. The developed MIMS method was applied to an analysis of air samples from an exhaust of a paintshop, where volatile organic solvents were used.
  • Keywords
    Air analysis , Aromatic compounds , Organochlorocompounds , Membrane inlet mass spectrometry
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1997
  • Journal title
    Analytica Chimica Acta
  • Record number

    1024668