    Lithium determination in artificial serum using flow injection systems with a selective solid-state tubular electrode based on NASICON membranes

  • Author/Authors

    M. Cretin، نويسنده , , Ll. Alerm، نويسنده , , J. Bartroli، نويسنده , , P. Fabry، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    8
  • From page
    7
  • To page
    14
  • Abstract
    The main objective of this work is the determination of lithium in the serum in the concentrations found in the plasma and the urine of patients being treated with lithium salts. Results of measurements under quiescent conditions and with flow injection analysis are compared. The solid-state tubular lithium sensor evaluated is based on conductive ceramics. A flow injection analysis (FIA) system using Li+-NASICON as the sensitive membrane and a BaCl2 0.1 M solution as carrier shows several advantages such as a good sampling frequency (20 samples h−1), a low detection limit (logarithmic range from 5 × 10−4 M) and a near-Nernstian sensitivity (57 mV/decade). Lithium measurements in artificial serum, using 0.1 M BaCl2 as a carrier solution, provided better results. Lithium determination in urine is possible because the usual concentration (10–30 mM) is in the logarithmic response range of the sensor. Concerning blood analysis (from 7 × 10−4 to 2 × 10−3 M in Li), the non-Nernstian response of the system in the lithium therapy concentration range does not seem to be a major problem because of the systematic calibration possibilities when using FIA instrumentation.
  • Keywords
    Artificial serum , Lithium analysis , How injection , Ion selective electrode , Li+-NASICON
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1997
  • Journal title
    Analytica Chimica Acta
  • Record number

    1024672