Title of article :
Evaluation of spectrophotometric methods for determination of orthophosphates by sequential injection analysis
Author/Authors :
A. Mu?oz، نويسنده , , Mas-Torres، Francisca نويسنده , , J.M. Estela، نويسنده , , V. Cerdà، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
9
From page :
21
To page :
29
Abstract :
Three methods for the determination of orthophosphate have been studied by using the same sequential injection analysis manifold, namely the vanadomolybdate, the Malachite Green and the molybdenum blue methods. The calibration graphs were linear up to 18.0, 0.4 and 4.0 mg P l−1, respectively. Relative standard deviations and detection limits for vanadomolybdate, Malachite Green and molybdenum blue were 2.1%, 18% and 1.7% and 0.15, 0.01 and 0.01 mg P l−1, respectively. The three methods were applied to the determination of phosphate in several water samples, both from natural and waste origin at a frequency of about 30 determinations per hour.
Keywords :
Phosphate , wastewater , Sequential injection analysis
Journal title :
Analytica Chimica Acta
Serial Year :
1997
Journal title :
Analytica Chimica Acta
Record number :
1024674
Link To Document :
بازگشت