• Title of article

    Lauryl sulphate as counter ion for construction of ion-selective membrane electrodes for moclobemide and disopyramide

  • Author/Authors

    R.I. Stefan، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    4
  • From page
    105
  • To page
    108
  • Abstract
    The construction and performance characteristics of two ion-selective membrane electrodes based on ion-pair complexes between lauryl sulphate and moclobemide (I) and disopyramide (II) are presented. The responses of the electrodes are near-Nernstian and the detection limits are 1 × 10−10 (I) and 7 × 10−8 mol 1−1 (II). The working pH ranges for the electrodes are 2.0–6.5 (I) and 2.0–8.0 (II). The selectivities of these electrodes were checked. Inorganic cations Na+, K+, Ca2+ do not interfere. Both membrane electrodes can be used for potentiometric titration of the active principle (I and II) of raw materials and drugs formulations.
  • Keywords
    Lauryl sulphate , Moclobemide-selective membrane electrode , Disopyramide-selective membrane electrode , Uniformity content test
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1997
  • Journal title
    Analytica Chimica Acta
  • Record number