Title of article :
Usefulness of a ruggedness test in the validation of flow injection analysis systems
Author/Authors :
Luis M.B.C ?lvares-Ribeiro، نويسنده , , Adélio A.S.C. Machado، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
7
From page :
195
To page :
201
Abstract :
The usefulness of a ruggedness test to evaluate the influence of small variations in the parameters fixed in the assemblage of a flow injection analysis (FIA) system on its response is presented. The purpose of the tested system was the determination of calcium in water with increased sensitivity by differential potentiometry involving two identical calcium ion selective electrodes. Youden and Steinerʹs ruggedness test was found to be a useful technique to ascertain that the optimized system worked properly.
Keywords :
Ruggedness test , FIA assembly , calcium , Differential potentiometry