Title of article :
Modeling and optimization studies for a sequential flow based bio-analytical system
Author/Authors :
Sameer Parab، نويسنده , , Bernard J. Van Wie، نويسنده , , Ian Byrnes، نويسنده , , Edgar J. Robles، نويسنده , , Bruce Weyrauch، نويسنده , , Thomas O. Tiffany، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
The clinical applications of the sequential flow technique are emphasized in this study by considering glucose determinations in a standard blood chemistry analysis. The sequential flow configuration is modeled to predict experimental behavior. A rigorous theoretical formulation of the model is presented incorporating temperature relationships for the hydrodynamic and kinetic parameters. A comparisons of model predictions with experiments show agreement satisfactory for determining optimal design strategies. To perform this analysis, factorial design results of model predictions are used to locate regions where a full set of model predictions should be made. Also, experimental results and theoretical predictions are used to determine the best temperature for glucose determinations.
Keywords :
Blood chemistry analyzer , Glucose , Sequential injection , Axial dispersion plug flow model , Flow injection analysis
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta