Title of article :
Selection of self-absorption free analytical lines of aluminum in glow discharge optical emission spectrometry
Author/Authors :
Yoshinari Inukai، نويسنده , , Yasuhiko Kaida، نويسنده , , Seiji Yasuda، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
7
From page :
267
To page :
273
Abstract :
To obtain a larger dynamic range in the glow discharge emission spectrometric determination of aluminum, the use of the ionic lines instead of the atomic resonance line was suggested. The ionic lines are essentially free from self-absorption leading to better linearity of analytical curves, while the atomic lines suffer from self-absorption to a certain extent. The ionic lines of aluminum are hardly emitted by the argon plasma; however, they can be excited by the neon-argon as well as the argon-helium mixed gas plasma. Therefore, the mixed plasma gases should be used. Calibration curves were measured for the different plasma gases and the concentration interval over which they were linear was estimated.
Keywords :
Atomic emission spectrometry , Self-absorption , Glow discharge , Ionic lines , Argon-helium plasma , Neon-argon plasma , Aluminum
Journal title :
Analytica Chimica Acta
Serial Year :
1997
Journal title :
Analytica Chimica Acta
Record number :
1025278
Link To Document :
بازگشت