• Title of article

    Use of a carbon paste modified electrode for the determination of 2-nitrophenol in a flow system by differential pulse voltammetry

  • Author/Authors

    Ignacio Naranjo-Rodriguez، نويسنده , , Juan A. Mu?oz Leyva، نويسنده , , JoséL. Hidalgo، نويسنده , , Hidalgo de Cisneros، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    7
  • From page
    167
  • To page
    173
  • Abstract
    The electrochemical reduction of 2-nitrophenol at a bentonite-modified carbon paste electrode by rapid scan differential pulse voltammetry in a flow system is studied, and a method for the direct voltammetric determination in sea water is proposed. A formic acid/sodium formate buffer solution of pH 4 was used as carrier. The potential range and the scan rate were selected according to the response of the analyte. 2-Nitrophenol showed a reduction potential within the cathodic range for the carbon electrode; a well-defined single peak was obtained. It is not necessary to remove the paste from the electrode after each new injection because the surface activity of the electrode remained constant at least during the time of a single complete analysis. A linear relationship between peak height and concentration was obtained in the range 0.05–5 mg l−1 with a detection limit of 0.02 mg l−1 and a relative standard deviation of 3.6%. Several chloro- and nitrophenols were tested to assess their influence on 2-nitrophenol signals.
  • Keywords
    Flow system , Carbon paste modified electrode , Differential pulse voltammetry , 2-Nitrophenol
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1997
  • Journal title
    Analytica Chimica Acta
  • Record number