• Title of article

    High selective determination of bromate in ozonized water by using ion chromatography with postcolumn derivatization equipped with reagent preparation device

  • Author/Authors

    Yoshinori Inoue، نويسنده , , Tetsushi Sakai، نويسنده , , Hiroki Kumagai، نويسنده , , Yuzuru Hanaoka، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    7
  • From page
    299
  • To page
    305
  • Abstract
    A sensitive and selective ion chromatographic determination method of bromate with postcolumn conversion into tribromide by hydrobromic acid has been developed. A high precision determination of bromate was performed by using the on-line hydrobromic acid generator employing a cation exchange hollow fiber. Under optimized conditions, the performance for bromate determination was obtained as given below. The detection limit (S/N=3) for a bromate injection of 0.1 ml was 0.35 μg l−1. The linear range was over 4 orders of magnitude, from 0.5 μg l−1 to 10 mg l−1. The relative standard deviation (RSD) for 1 μg l−1 of bromate solution was 4.6% (n=10). The postcolumn derivatization systems combined with suppressed conductometric detection was applied to the simultaneous determination of bromate and common anions, such as fluoride, chloride, nitrite, bromide, phosphate, sulfate, and nitrate, in drinking water.
  • Keywords
    Bromate , Ozonized water , Ion chromatography , Postcolumn derivatization , Tribromide
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1997
  • Journal title
    Analytica Chimica Acta
  • Record number