Title of article
Analysis of some non-linear effects in quantitative electron paramagnetic resonance spectrometry. Non-linear effects due to the use of cavities with TE102 and TM110 modes
Author/Authors
Nicola D. Yordanov، نويسنده , , Biljana Genova، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
5
From page
99
To page
103
Abstract
The non-linear effects between the electron paramagnetic resonance (EPR) signal intensity and sample shape (length and diameter) and its positioning in the most commonly accepted cavities operating in TE102 or TM110 modes are discussed in detail. These effects are only due to the physical properties of the cavities. It is obvious from both theoretical considerations and experimental results that several precautions must be taken with respect to the physical parameters, like internal diameter of the sample tubes and length of the unknown sample and standard on planning quantitative EPR estimations.
Keywords
electron paramagnetic resonance , Cavities , Non-linear effects
Journal title
Analytica Chimica Acta
Serial Year
1997
Journal title
Analytica Chimica Acta
Record number
1025388