Title of article :
Dynamic reaction cell ICP-MS for determination of total As, Cr, Se and V in complex matrices: Still a challenge? A review
Author/Authors :
Sonia D’Ilio، نويسنده , , Nicola Violante، نويسنده , , Costanza Majorani، نويسنده , , Francesco Petrucci، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
Mass interferences, caused by atomic or polyatomic species and having the same mass/charge ratio of the analyte, can be a severe limit for a reliable assay of trace and ultratrace elements by ICP-MS. The DRC™ technology uses a reaction gas to overcome these interferences. Reactions of charge exchange, atom transfer, adduct formation, condensation and analyte association/condensation are the main mechanisms. Interfering ions tend to react with the gas exothermally, while, the analyte reacts endothermally.
Keywords :
Surface-enhanced Raman scattering , Competitive , Label , Immunoassay , Clenbuterol
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta