• Title of article

    High temperature liquid chromatography hyphenated with ESI-MS and ICP-MS detection for the structural characterization and quantification of halogen containing drug metabolites Original Research Article

  • Author/Authors

    Jon S.B. de Vlieger، نويسنده , , Mark J.N. Giezen، نويسنده , , David Falck، نويسنده , , Cornelis Tump، نويسنده , , Fred van Heuveln، نويسنده , , Martin Giera، نويسنده , , Jeroen Kool، نويسنده , , Henk Lingeman، نويسنده , , Jaap Wieling، نويسنده , , Maarten Honing، نويسنده , , Hubertus Irth، نويسنده , , Wilfried M.A. Niessen، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    8
  • From page
    69
  • To page
    76
  • Abstract
    In this paper we describe the hyphenation of high temperature liquid chromatography with ICP-MS and ESI-MS for the characterization of halogen containing drug metabolites. The use of temperature gradients up to 200 °C enabled the separation of metabolites with low organic modifier content. This specific property allowed the use of detection methods that suffer from (significant) changes in analyte response factors as a function of the organic modifier content such as ICP-MS. Metabolites of two kinase inhibitors (SB-203580-Iodo and MAPK inhibitor VIII) produced by bacterial cytochrome P450 BM3 mutants and human liver microsomes were identified based on high resolution MSn data. Quantification was done using their normalized and elemental specific response in the ICP-MS. The importance of these kinds of quantification strategies is stressed by the observation that the difference of the position of one oxygen atom in a structure can greatly affect its response in ESI-MS and UV detection.
  • Keywords
    DRC-ICP-MS , Interferences suppression , Reaction gases , Trace elements
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2011
  • Journal title
    Analytica Chimica Acta
  • Record number