Title of article :
An improved method for determination of lanthanoids in environmental samples by inductively coupled plasma mass spectrometry with high matrix introduction system Original Research Article
Author/Authors :
Valbona Celo a، نويسنده , , Ewa Dabek-Zlotorzynska، نويسنده , , Jiujiang Zhao a، نويسنده , , Irina Okonskaia، نويسنده , , David Bowman، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
The capability of ICP-MS equipped with the high matrix introduction system (HMI) for accurate analysis of lanthanoids in environmental samples was investigated. Compared to the conventional operation, the amounts of oxide and hydroxide molecular species formed in the plasma were reduced by up to 5 times. The relative yields of oxides did not exceed 0.02% for BaO+ species and were as low as 0.3% for lanthanoids with the highest oxide-formation rates (LaO+, CeO+, PrO+ and NdO+). Hydroxide formation was less than 0.02% when HMI system was used. In addition, two digestion procedures were evaluated by the analysis of standard reference materials (SRMs) of different matrices. The digestion efficiency and limits of detection (LOD) were improved when samples were digested with a mixture of HNO3/H2O2/HCl/HF, and HF was removed by evaporation in presence of concentrated HCl. Using this procedure and HMI-ICP-MS analysis, LOD ranged from 0.1 μg kg−1 to 6 μg kg−1. Recoveries ranged from 85 to 115% for La to Ho and from 75 to 85% for the other lanthanoids. Relative standard deviations for replicate analysis of SRMs were less than 10%.
Keywords :
Microwave digestion , Environmental samples , Lanthanoids , High matrix introduction system (HMI) , Inductively coupled plasma mass spectrometry (ICP-MS)
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta