• Title of article

    An automated screening system for benzodiazepines in human urine Original Research Article

  • Author/Authors

    D Gambart، نويسنده , , S C?rdenas، نويسنده , , M Gallego، نويسنده , , M Valc?rcel، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    10
  • From page
    93
  • To page
    102
  • Abstract
    A rapid flow system for determining the presence or absence of benzodiazepines in human urine has been developed and optimized. The system is based on the diazotization of the analytes in a pre-hydrolyzed samples at pH<1, followed by the formation of dyes with 1-naphthol at pH>12. The dyes are retained and washed on a XAD-2 minicolumn, and immediately eluted with 3 M NaOH in ethanol/water. A simple photometer is used at λ=550 or 600 nm to monitor the overall analyte content. The reliability of the proposed method for the benzodiazepines is demonstrated in simple chemometric terms, using four concentration levels (between 0.5 and 2 times the detection limit) and oxazepam as model benzodiazepine. The method is highly selective and a sensitive enough for detecting benzodiazepines in urine.
  • Keywords
    Screening system , False positives and negatives , Urine , Benzodiazepines
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1998
  • Journal title
    Analytica Chimica Acta
  • Record number

    1026898