Title of article :
Electrothermic factors optimization in electrothermal atomic absorption spectrometry via an optimal experimental design matrix Original Research Article
Author/Authors :
Hélène Le Garrec، نويسنده , , Philippe Giamarchi، نويسنده , , Jean Yves Cabon، نويسنده , , Alain Le Bihan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
12
From page :
59
To page :
70
Abstract :
A new approach is proposed to optimize some instrumental electrothermic parameters, i.e. injected volume, calcination duration, calcination and atomization temperatures, gas flow, in electrothermal atomic absorption spectrometry. It is based on the monitoring of the detection limit estimation variation by using an unconventional experimental design well adapted to a particular experimental domain. This method was tested on Cd, Mn and Cu in sea water. By comparison with the default operating conditions, the limits of detection were improved in some cases and always obtained on carrying out less experiments than usual.
Keywords :
Detection limit estimation , Electrothermal atomic absorption spectrometry , Optimal experimental design , Electrothermic factors optimization