Title of article :
Linear calibration function for optical oxygen sensors based on quenching of ruthenium fluorescence Original Research Article
Author/Authors :
Tsung-Han Chuang، نويسنده , , Mark A Arnold، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
A mathematical model is derived from the Stern–Volmer equation as a practical calibration function for various optical oxygen sensors based on quenching of ruthenium fluorescence. The model is simple and flexible, and is ideal for computing oxygen partial pressure from 0 to 760 Torr in gaseous samples, and 0 to 200 Torr in aqueous samples. Feasibility of this model for dissolved oxygen calibration is demonstrated by applying the model to data acquired from a set of three radioluminescence-based optical oxygen sensors. The model is applicable to various sensor designs and adequately accounts for stray radiation and other sources of non-quenchable light.
Keywords :
Radioluminescent oxygen sensors , Oxygen sensors , Ruthenium fluorescence quenching , Optical sensors
Journal title :
Analytica Chimica Acta
Journal title :