Title of article :
Potentiometric enzyme electrode in a flow injection system for the determination of urea in human serum samples Original Research Article
Author/Authors :
Izabela Wa?cerz، نويسنده , , Stanis?aw G??b، نويسنده , , Robert Koncki، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
9
From page :
129
To page :
137
Abstract :
A urea biosensor with urease covalently bound directly to the membrane of the ammonium ion selective electrode has been developed. The enzyme electrode was tested in single and double channel flow injection analysis (FIA) systems. Optimised measurement conditions allowed determination of urea within the linear range of 0.5–8 mM in the single channel and of 1–10 mM in the double channel FIA system. The response time of the biosensor was short enabling frequent sample injections (every 1.5–3 min). Long term stability of the enzyme electrode allows application of the same biosensor in the FIA systems for up to two months. Under optimised conditions the determination of urea is possible at physiological and pathological levels. The results of urea determination in real, human serum samples were in good agreement with those obtained by routine clinical analysis methods.
Keywords :
Flow injection analysis , Serum , Enzyme electrode , Urea
Journal title :
Analytica Chimica Acta
Serial Year :
1998
Journal title :
Analytica Chimica Acta
Record number :
1026997
Link To Document :
بازگشت