Title of article :
Potentiometric biosensor for lysine analysis based on a chemically immobilized lysine oxidase membrane Original Research Article
Author/Authors :
Javier Saurina، نويسنده , , Santiago Hern?ndez-Cassou، نويسنده , , Esteve Fàbregas، نويسنده , , Salvador Alegret، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
8
From page :
49
To page :
56
Abstract :
An enzymatic lysine electrode is developed for the potentiometric analysis of lysine (Lys). This lysine biosensor consists of a chemically immobilized lysine oxidase membrane attached to an all-solid-state ammonium electrode. In this system, the enzymatic degradation of lysine produces ammonia, which is then detected with the ammonium electrode. The influence of immobilization conditions on the potentiometric response and the lifetime of the electrode was studied. Under optimal conditions, the relationship between the potentiometric response and the log [Lys] was linear between 3×10−5 and 10−3 M, the sensitivity was 50 mV/decade, the repeatability 1.4%, and the response time t95% was 15 s.
Keywords :
Potentiometric lysine biosensor , All-solid-state ammonium electrode , Lysine analysis , Immobilized lysine oxidase membranes
Journal title :
Analytica Chimica Acta
Serial Year :
1998
Journal title :
Analytica Chimica Acta
Record number :
1027048
Link To Document :
بازگشت