• Title of article

    Direct spectrophotometric determination of chromium by microwave-oven induced flow-injection analysis Original Research Article

  • Author/Authors

    Xiaoyuan Wu، نويسنده , , Hui Zhao، نويسنده , , Xingguo Chen، نويسنده , , Zhide Hu، نويسنده , , Zhengfeng Zhao، نويسنده , , Martin Hooper، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    6
  • From page
    61
  • To page
    66
  • Abstract
    Two sensitive flow-injection methods have been developed for the determination of traces of Cr(VI), based on the very efficient microwave-accelerated reactions of Cr2O2−7 with dibromocarboxyarsenazo (DBM-AsA, in System 1) and dibromo-o-carborylchlorophonazo (DBOK-CPA, in System 2). The reactions are followed spectrophotometrically by measuring the decrease in the absorbance of DBM-AsA at 535 nm or DBOK-CPA at 556 nm. The sampling frequency for both systems is 40 h−1. The calibration graphs are linear in the range between 1.62 and 27.0 μg ml−1, and 0.40 and 10.80 μg ml−1 for Systems 1 and 2, respectively. The respective detection limits are 0.087 and 0.10 μg ml−1. The methods were applied to determining traces of chromium in a steel sample and electroplating solutions with satisfactory results.
  • Keywords
    Microwave , Spectrophotometric flow-injection , Chromium(VI) , Dibromo-o-carborylchlorophonazo , Dibromocarboxyarsenazo
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1998
  • Journal title
    Analytica Chimica Acta
  • Record number

    1027152