• Title of article

    Polyethyleneimine-coated reticulated vitreous carbon electrode with immobilized enzymes as a substrate detector Original Research Article

  • Author/Authors

    Xiaojing Tang، نويسنده , , Bin Xie، نويسنده , , Per-Olof Larsson، نويسنده , , Bengt Danielsson، نويسنده , , Masoud Khayyami، نويسنده , , Gillis Johansson، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    6
  • From page
    185
  • To page
    190
  • Abstract
    Polyethyleneimine (PEI) was covalently coupled to carbodiimide-activated reticulated vitreous carbon (RVC). The PEI-coated RVC was activated with glutaraldehyde, and glycerol dehydrogenase and diaphorase were then immobilized. The PEI-coated RVC with immobilized enzymes functioned both as an enzyme reactor and a working electrode in an amperometric detection system where NAD+/NADH was recycled by the immobilized enzymes. The coated RVC electrode showed good properties compared to uncoated RVC, such as a long lifetime and a constant response to a series of injections in a flow-injection system, resulting in a relative standard deviation of 1.4%. The calibration graph was linear from the detection limit 0.1 μM to 2 mM NADH in the absence of recycling.
  • Keywords
    Polyethyleneimine , Reticulated vitreous carbon , Glycerol dehydrogenase , Diaphorase , Flow injection
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1998
  • Journal title
    Analytica Chimica Acta
  • Record number

    1027163