Abstract
A novel electrochemical immunosensor based on the principles of liposome signal amplification and immunochromatography has been developed for the determination of theophylline. The immunosensor is composed of two major parts including a screen-printed electrode and an immunochromatographic nitrocellulose membrane strip. On the membrane, anti-theophylline antibody is immobilized in an antibody competition zone and hexacyanoferrate(II)-loaded liposomes are immobilized in a signal generation zone. When a theophylline sample solution is applied to the immunosensor pre-loaded with theophylline–melittin conjugate in sample loading zone, the theophylline and theophylline–melittin conjugate migrate through the anti-theophylline antibody zone, where competitive binding occurs. Unbound theophylline–melittin conjugate further migrates into the signal generation zone, where it disrupts the liposomes to release the electroactive hexacyanoferrate(II) which is then detected amperometrically. The current produced is directly proportional to the concentration of hexacyanoferrate(II) which, in turn, can be related to the concentration of free analyte in the sample. The detection limit of 5 μg ml−1 enables the present immunosensor to be used to monitor theophylline over the clinically relevant ranges (10–20 μg ml−1) with a one-step assay within 20 min.
Keywords
Immunochromatographic membrane , Immunosensor , Theophylline , Liposome , Field-portable sensor
Journal title
Analytica Chimica Acta
Serial Year
1999
Journal title
Analytica Chimica Acta
Record number
1027342
Link To Document