Title of article :
Instrumentation, standards, and procedures used at the National Research Council of Canada for high-accuracy fluorescence measurements Original Research Article
Author/Authors :
Joanne C. Zwinkels، نويسنده , , François Gauthier، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
This paper describes the two-monochromator reference spectrofluorimeter, calibration standards, and procedures used at the National Research Council of Canada for performing high-accuracy fluorescence measurements over the spectral range 250–1050 nm. The basic measurement principles of the two-monochromator method and the calibration procedures used to determine the instrument spectral characteristics are described in detail. The instrumental set-ups, data acquisition procedures, and representative results for the measurement of reflected and bispectral luminescent radiance factors and calculation of total radiance factors are graphically illustrated in a step-by-step fashion for a typical red fluorescent material. The measurement performance of this instrument is demonstrated for several applications, including customer-supplied source data, paper, paint and textile samples.
Keywords :
fluorescence , calibration , Standards , NRC , Colorimetry , Spectrofluorimeter
Journal title :
Analytica Chimica Acta