Two-way, unfolded three-way and three-mode partial least squares calibration of diode array HPLC chromatograms for the quantitation of low-level pharmaceutical impurities Original Research Article
Author/Authors :
Konstantinos D. Zissis، نويسنده , , Richard G. Brereton، نويسنده , , Samantha Dunkerley، نويسنده , , Richard E.A. Escott، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
A series of two-way diode array chromatograms were recorded at 0.1–0.5% and 1–5% of 3-hydroxypyridine impurity co-eluting with 2-hydroxypyridine, recorded at pH 4.9 (good resolution) and 5.0 (poor resolution). Four methods for PLS calibration, namely, using summed spectral profiles, summed elution profiles, unfolded three-way PLS and true three-way PLS were applied to the datasets, both using autopredictions and cross-validation. It was found that it was possible to accurately quantify low levels of impurities. Three-way methods often performed worse than two-way methods using the summed spectral profile probably due to irreproducibility of elution times.
Keywords :
HPLC , Three-mode calibration , partial least squares , Impurity monitoring