Title of article :
Determination of gallium by graphite furnace atomic absorption spectrometry with combined use of a tungsten-coated Lʹvov platform tube and a chemical modification technique Original Research Article
Author/Authors :
Dianjie Ma، نويسنده , , Yasuaki Okamoto، نويسنده , , Takahiro Kumamaru، نويسنده , , Etsuro Iwamoto، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
Tungsten-coated non-pyrolytic graphite (NPG), pyrolytic graphite (PG) and pyrolytic Lʹvov platform graphite (PPG) tubes were prepared, and their analytical performances were compared. The coating process simply involved injecting 100 μl of a sodium tungstate solution (0.01 mol l−1) into each graphite tube, followed by heating according to a temperature programme similar to an atomisation cycle for the determination of gallium. This procedure for coating was repeated at least 12, 25, and 7 times towards NPG, PG, and PPG tubes, respectively. Among these tubes, the tungsten-coated PPG tube showed excellent performance for the determination of gallium. By combined use of a chemical modifier such as aluminium(III) or nickel(II) a detection limit (3σ) of 6 pg and sensitivity (1% absorption) of 3–4 pg were achieved. The practical potential of the proposed technique was demonstrated for the determination of gallium in several samples of alloys and fresh water.
Keywords :
Tungsten-coated Lיvov platform tube , Graphite furnace atomic absorption spectrometry , Chemical modification , Gallium
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta