Title of article :
Determination of trimethylselenonium ion by flow injection hydride generation atomic absorption spectrometry Original Research Article
Author/Authors :
Amit Chatterjee، نويسنده , , Y Shibata، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
6
From page :
273
To page :
278
Abstract :
A simple direct flow injection hydride generation atomic absorption spectrometric method was developed using sodium tetrahydroborate and hydrochloric acid for the determination of trimethylselenonium (TmSe) iodide which excluded any chemical pretreatment prior to hydride generation. The detection limit (2σ of blank) of the method was 1.0 ng ml−1 selenium and the sensitivity was 0.15 ng ml−1. The calibration graph was linear up to 30 ng ml−1. The recoveries from tap water and urine varied with dilution, being 89–95% from tap water at 1 : 11 dilution, and 50–58% from urine at 1 : 50 dilution.
Keywords :
Selenium , Hydride generation , Trimethylselenonium ion , Flow injection hydride generation atomic absorption spectrometry
Journal title :
Analytica Chimica Acta
Serial Year :
1999
Journal title :
Analytica Chimica Acta
Record number :
1027992
Link To Document :
بازگشت