Title of article :
Comparison of quantification strategies for one-point standard addition calibration: The homoscedastic case Original Research Article
Author/Authors :
Richard J.C. Brown، نويسنده , , Thomas P.S. Gillam، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
4
From page :
108
To page :
111
Abstract :
Two quantification strategies for one-point standard addition calibration have been compared mathematically. One strategy involved the extrapolation of measurement points to their intercept with the x-axis to determine the analyte content in the unknown sample, and the other strategy is based upon direct calculation of the analyte content in the unknown sample using the instrumental responses obtained during measurement. The cases of both conventional standard addition calibration (C-SAC) and sequential standard addition calibration (S-SAC) have been considered. The homoscedastic situation has been considered, where the absolute precision of the instrumental response is constant. It has been determined that the precision ratio of the two strategies is dependent on surprisingly simple parameters: such as the sample to standard mass ratio (for C-SAC) and the analyte content ratio (for S-SAC).
Keywords :
Standard addition , calibration , Precision , extrapolation , Comparison
Journal title :
Analytica Chimica Acta
Serial Year :
2012
Journal title :
Analytica Chimica Acta
Record number :
1028137
Link To Document :
بازگشت