Development of a novel label-free amperometric immunosensor for the detection of okadaic acid Original Research Article
Author/Authors :
Akhtar Hayat، نويسنده , , Lise Barthelmebs، نويسنده , , Audrey Sassolas، نويسنده , , Jean-Louis Marty، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
6
From page :
92
To page :
97
Abstract :
Okadaic acid (OA), a lipophilic phycotoxin is mainly produced by toxigenic dinoflagellates. The need to develop high performing methods for OA analysis able to improve the traditional ones is evident. In this work, a novel experimental methodology for label-free detection of OA was developed. Protein G magnetic beads (protein-G-MBs) modified gold electrode was used to immobilize anti-OA monoclonal antibody (anti-OA-MAb). Preliminary, colorimetric tests were performed in order to validate protein-G-MBs and anti-OA-MAb reaction. Electrochemical detection was carried out by differential pulse voltammetry in ferri/ferrocyanide solution. The limit of detection value obtained (0.5 μg L−1) validated the developed electrochemical immunosensor as a promising tool for routine use. The matrix effect and the recovery rate were also assessed with real samples, showing a good percentage of recovery.
Keywords :
Mussel extract , Okadaic acid , Label-free immunosensor , Protein G magnetic beads , Amperometric immunosensor