Title of article :
Novel sample preparation technique with needle-type micro-extraction device for volatile organic compounds in indoor air samples Original Research Article
Author/Authors :
Ikuo Ueta، نويسنده , , Ayako Mizuguchi، نويسنده , , Koji Fujimura، نويسنده , , Susumu Kawakubo، نويسنده , , Yoshihiro Saito، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Abstract :
A novel needle-type sample preparation device was developed for the effective preconcentration of volatile organic compounds (VOCs) in indoor air before gas chromatography–mass spectrometry (GC–MS) analysis. To develop a device for extracting a wide range of VOCs typically found in indoor air, several types of particulate sorbents were tested as the extraction medium in the needle-type extraction device. To determine the content of these VOCs, air samples were collected for 30 min with the packed sorbent(s) in the extraction needle, and the extracted VOCs were thermally desorbed in a GC injection port by the direct insertion of the needle. A double-bed sorbent consisting of a needle packed with divinylbenzene and activated carbon particles exhibited excellent extraction and desorption performance and adequate extraction capacity for all the investigated VOCs. The results also clearly demonstrated that the proposed sample preparation method is a more rapid, simpler extraction/desorption technique than traditional sample preparation methods.
Keywords :
Gas chromatography–mass spectrometry , Needle , Volatile organic compound , Extraction , Indoor air