Title of article :
Non-linear calibration in single point flow titration of protolytes: A comparison of model options by using Delta Test Original Research Article
Author/Authors :
Fernando Albert?s، نويسنده , , Isel Cortés، نويسنده , , Manuel Alvarez، نويسنده , , Victor Cerda، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
17
From page :
221
To page :
237
Abstract :
The complete description of the non-linear absorbance/concentration relationship obtained for systems used in single point photometric flow injection titrations of acids and bases is presented. The applicability of different models for fitting calibration data, as well as the more frequently used data processing options, are investigated under normal and restraining conditions by using experimental and simulated data sets. Theoretical and empirical models are examined in order to choose the simplest and the more efficient ones to provide good recoveries. The comparison among models is based on a novel statistical criterion (Delta Test) for contrasting the prediction efficiency by detecting systematic errors in estimated concentrations.
Keywords :
Bases , Curved calibration , Regression models , Efficiency of analytical functions , Goodness of fit , FIA , Acids , Photometric flow titration
Journal title :
Analytica Chimica Acta
Serial Year :
2000
Journal title :
Analytica Chimica Acta
Record number :
1029183
Link To Document :