Title of article
Disposable biosensor based on cathodic electrochemiluminescence of tris(2,2-bipyridine)ruthenium(II) for uric acid determination Original Research Article
Author/Authors
J. Ballesta-Claver، نويسنده , , R. Rodr?guez-G?mez، نويسنده , , L.F. Capit?n-Vallvey، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2013
Pages
8
From page
153
To page
160
Abstract
A new method for uric acid (UA) determination based on the quenching of the cathodic ECL of the tris(2,2-bipyridine)ruthenium(II)–uricase system is described. The biosensor is based on a double-layer design containing first tris(2,2-bipyridine)ruthenium(II) (Ru(bpy)32+) electrochemically immobilized on graphite screen-printed cells and uricase in chitosan as a second layer. The uric acid biosensing is based on the ECL quenching produced by uric acid over the cathodic ECL caused by immobilized Ru(bpy)32+ in the presence of uricase. The use of a −1.1 V pulse for 1 s with a dwelling time of 10 s makes it possible to estimate the initial enzymatic rate, which is used as the analytical signal. The Stern–Volmer type calibration function shows a dynamic range from 1.0 × 10−5 to 1.0 × 10−3 M with a limit of detection of 3.1 × 10−6 M and an accuracy of 13.6% (1.0 × 10−4 M, n = 5) as relative standard deviation. Satisfactory results were obtained for urine samples, creating an affordable alternative for uric acid determination.
Keywords
Disposable biosensor , Cathodic electrochemiluminescence , Immobilized Ru(bpy)32+ , Uric acid determination
Journal title
Analytica Chimica Acta
Serial Year
2013
Journal title
Analytica Chimica Acta
Record number
1029313
Link To Document