Title of article :
Efficient use of pure component and interferent spectra in multivariate calibration Original Research Article
Author/Authors :
Sandeep Sharma، نويسنده , , Mohammad Goodarzi، نويسنده , , Laure Wynants، نويسنده , , Herman Ramon، نويسنده , , Wouter Saeys، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
9
From page :
15
To page :
23
Abstract :
Partial Least Squares (PLS) is by far the most popular regression method for building multivariate calibration models for spectroscopic data. However, the success of the conventional PLS approach depends on the availability of a ‘representative data set’ as the model needs to be trained for all expected variation at the prediction stage. When the concentration of the known interferents and their correlation with the analyte of interest change in a fashion which is not covered in the calibration set, the predictive performance of inverse calibration approaches such as conventional PLS can deteriorate. This underscores the need for calibration methods that are capable of building multivariate calibration models which can be robustified against the unexpected variation in the concentrations and the correlations of the known interferents in the test set. Several methods incorporating ‘a priori’ information such as pure component spectra of the analyte of interest and/or the known interferents have been proposed to build more robust calibration models. In the present study, four such calibration techniques have been benchmarked on two data sets with respect to their predictive ability and robustness: Net Analyte Preprocessing (NAP), Improved Direct Calibration (IDC), Science Based Calibration (SBC) and Augmented Classical Least Squares (ACLS) Calibration. For both data sets, the alternative calibration techniques were found to give good prediction performance even when the interferent structure in the test set was different from the one in the calibration set. The best results were obtained by the ACLS model incorporating both the pure component spectra of the analyte of interest and the interferents, resulting in a reduction of the RMSEP by a factor 3 compared to conventional PLS for the situation when the test set had a different interferent structure than the one in the calibration set.
Keywords :
Science Based Calibration , Augmented Classical Least Squares , Robust calibration , Glucose , Net Analyte Preprocessing , Improved Direct Calibration
Journal title :
Analytica Chimica Acta
Serial Year :
2013
Journal title :
Analytica Chimica Acta
Record number :
1029404
Link To Document :
بازگشت