Title of article :
A flow injection analysis system for monitoring silver (I) ion and iodine residuals in recycled water from recovery systems used for spaceflight Original Research Article
Author/Authors :
Jill P. Williamson، نويسنده , , Gary L. Emmert، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
7
From page :
72
To page :
78
Abstract :
A laboratory-built flow injection analyzer is reported for monitoring the drinking water disinfectants silver (I) ion and iodine in water produced from NASAʹs water recovery system. This analyzer uses spectrophotometric detection with a custom made 10 cm optical flow cell. Optimization and interference studies are discussed for the silver (I) ion configuration. Subsequent results using the silver (I) configuration with minor modifications and alternative reagents gave promising results for iodine determinations as well. The estimated MDL values for Ag+ and I2 are 52 μg L−1 Ag+ and 2 μg L−1 I2; the mean percent recoveries were 104% and 96.2% for Ag+ and I2 respectfully; and percent relative standard deviations were estimated at 1.4% for Ag+ and 5.7% for I2. The agreement of this potentially multifunctional analyzer to reference methods for each respective water disinfectant is measured using Bland–Altman analysis as well as more traditional estimates.
Keywords :
Flow analysis , Silver ion , Iodine , Disinfectant
Journal title :
Analytica Chimica Acta
Serial Year :
2013
Journal title :
Analytica Chimica Acta
Record number :
1029617
Link To Document :
بازگشت